is required to have, unless exempt under section 360j (g) of this title, an approval under this section of an application for premarket approval or, as applicable, an approval under subsection (c)(2) of this section of a report seeking premarket approval.
the Secretary shall by regulation, promulgated in accordance with this subsection, require that such device have an approval under this section of an application for premarket approval.
refer such application to the appropriate panel under section 360c of this title for study and for submission (within such period as he may establish) of a report and recommendation respecting approval of the application, together with all underlying data and the reasons or basis for the recommendation. Where appropriate, the Secretary shall ensure that such panel includes, or consults with, one or more pediatric experts.
In making the determination whether to approve or deny the application, the Secretary shall rely on the conditions of use included in the proposed labeling as the basis for determining whether or not there is a reasonable assurance of safety and effectiveness, if the proposed labeling is neither false nor misleading. In determining whether or not such labeling is false or misleading, the Secretary shall fairly evaluate all material facts pertinent to the proposed labeling.
Any denial of an application shall, insofar as the Secretary determines to be practicable, be accompanied by a statement informing the applicant of the measures required to place such application in approvable form (which measures may include further research by the applicant in accordance with one or more protocols prescribed by the Secretary).
that was not described as a deficiency in the written description provided by the Secretary under clause (ii).
the Secretary shall, unless he finds the petition to be without good cause or unless a petition for review of such order has been submitted under paragraph (2), hold a hearing, in accordance with section 554 of title 5, on the order. The panel or panels which considered the application, protocol, or device subject to such order shall designate a member to appear and testify at any such hearing upon request of the Secretary, the petitioner, or the officer conducting the hearing, but this requirement does not preclude any other member of the panel or panels from appearing and testifying at any such hearing. Upon completion of such hearing and after considering the record established in such hearing, the Secretary shall issue an order either affirming the order subject to the hearing or reversing such order and, as appropriate, approving or denying approval of the application, reinstating the applications approval, approving the protocol, or placing in effect a notice of completion.
the Secretary shall refer the application or protocol subject to the order and the basis for the order to an advisory committee of experts established pursuant to subparagraph (B) for a report and recommendation with respect to the order. The advisory committee shall, after independent study of the data and information furnished to it by the Secretary and other data and information before it, submit to the Secretary a report and recommendation, together with all underlying data and information and a statement of the reasons or basis for the recommendation. A copy of such report shall be promptly supplied by the Secretary to any person who petitioned for such referral to the advisory committee.
revising the classification of the device so that the device is classified into class I or class II, unless the regulation requires the device to remain in class III. In determining whether to revise the classification of a device or to require a device to remain in class III, the Secretary shall apply the criteria set forth in section 360c (a) of this title. Before the publication of a regulation requiring a device to remain in class III or revising its classification, the Secretary shall publish a proposed regulation respecting the classification of a device under this paragraph and provide reasonable opportunity for the submission of comments on any such regulation. No regulation requiring a device to remain in class III or revising its classification may take effect before the expiration of 90 days from the date of its publication in the Federal Register as a proposed regulation.